Protocols below are for use with Pasco probes. Click here
for protocols to use the Vernier temperature probe.
Materials:
PasPort Xplorer |
|
PasPort temperature sensor |
PasPort temperature probe |
USB cable |
DataStudio software |
Ice water |
Boiling water |
Steps for calibrating the temperature sensor:
1. Connect the temperature probe to the temperature sensor.
2. Connect the temperature sensor to the Pasport Xplorer device.
3. Connect the PasPort Xplorer device to the computer with the USB cable.
4. Open the DataStudio software and select “Launch DataStudio.”
(or "Create an Experiment")
5. Click on the “Setup” button located near the top of the
screen.
6. Click on the “Calibrate” button in the setup window.
7. Place the temperature sensor probe into the ice water bath.
8. Enter the 0.0 C for “Point 1” and click the “Set”
button when the temperature reading stabilizes. (Note: The temperature
reading value need not be the set value.)
9. Removed the temperature sensor probe from the ice water bath.
10. Repeat steps 7 – 8 with the boiling water, entering this 100.0
C for “Point 2.”
11. Press the "Ok" button and the sensor should be calibrated.